Field replaceable units: Replace main unit parts
7-46 Dash 3000/4000/5000 2000966-542D
5. Unscrew the thumb screw anchoring the writer cable to the CPU
assembly and disconnect the flex cable from the CPU/battery housing
assembly.
6. Remove the three screws anchoring the battery door assembly to the
rear housing. Remove the battery door assembly.
7. If replacing the battery door only, install the new door and
reassemble the patient monitor. Otherwise, continue with the next
step.
8. Remove the four panhead screws anchoring the CPU/battery housing
assembly to the frame. These screws are attached at the bottom of
the unit.
Screws holding battery door.
581B
895A
517A