EasyManuals Logo
Home>GE>Monitor>Dash 3000

GE Dash 3000 Service Manual

GE Dash 3000
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #260 background imageLoading...
Page #260 background image
Electromagnetic compatibility (EMC): Electromagnetic compatibility (EMC)
A-4 Dash 3000/4000/5000 2000966-542D
NOTE
U
t
is the AC mains voltage prior to application of the test level.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV contact ± 6 kV contact Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
± 8 kV air ± 4 kV air
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for power supply lines ± 2 kV for power supply
lines
Mains power should be that of a typical
commercial or hospital environment.
±1 kV for input/output lines ±1 kV for input/output lines
Surge
IEC 61000-4-5
± 1 kV differential mode ± 1 kV differential mode Mains power should be that of a typical
commercial or hospital environment.
± 2 kV common mode ± 2 kV common mode
<5% U
t
(>95% dip in U
t
)
for 0.5 cycles
<5% U
t
(>95% dip in U
t
) for
0.5 cycles
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<40% U
t
(>60% dip in U
t
)
for 5 cycles
<40% U
t
(>60% dip in U
t
) for
5 cycles
Mains power should be that of a typical
commercial or hospital environment. If the user
of the equipment requires continued operation
during power mains interruptions, it is
recommended that the equipment be powered
from an uninterruptible power supply or a battery.
<70% U
t
(>30% dip in U
t
)
for 25 cycles
<70% U
t
(>30% dip in U
t
) for
25 cycles
<5% U
t
(>95% dip in U
t
)
for 5 s
<5% U
t
(>95% dip in U
t
) for
5 s
Power Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.

Table of Contents

Other manuals for GE Dash 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 3000 and is the answer not in the manual?

GE Dash 3000 Specifications

General IconGeneral
BrandGE
ModelDash 3000
CategoryMonitor
LanguageEnglish

Related product manuals