Configuration: Service menus
4-8 Dash 3000/4000/5000 2000966-542D
MONITOR SETTINGS (cont.) CONFIGURE WIRELESS LAN
Enable and configure the wireless LAN option. (Refer
to “Confirm or configure wireless LAN” on page 4-14
for detailed procedures.)
WIRELESS LAN STATUS
Displays wireless LAN status.
STORE DEFAULTS FOR NETWORK TRANSFER
Saves the patient monitor defaults for transfer to
another patient monitor with the same software
version, software package, patient monitor type, and
country code. It stores all patient monitor defaults,
(including WLAN configuration) custom default
names, and 12 SL location and site.
SET LOCATION ID
Enter the location identification for the MUSE system.
SET SITE NUMBER
Enter a site number if your institution has multiple
buildings. A MUSE system can be used as a site
number.
COPY UNIT DEFAULTS Use to copy patient monitor defaults from one patient
monitor to another patient monitor. Refer to
“Configuration” on page 4-1 for detailed procedures.
PRINT TEST PATTERN Allows you to run a graph test pattern. The choices are
Start and Stop.
TIME AND DATE Allows changes to the time and date and may affect the
time and date for the entire monitoring network. (Refer to
“Set time and date” on page 4-23 for detailed
procedures.)
WARNING
Loss of patient history. This menu should rarely be used
because patient histories will be lost.
Service Mode menu options (Continued)
Menu option Description

Table of Contents

Other manuals for GE Dash 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 3000 and is the answer not in the manual?

GE Dash 3000 Specifications

General IconGeneral
BrandGE
ModelDash 3000
CategoryMonitor
LanguageEnglish

Related product manuals