Configuration: Advanced user procedures
2000966-542D Dash 3000/4000/5000 4-25
Store monitor defaults for transfer
To transfer monitor defaults, they must be set, then stored on the patient
monitor that is used as the server.
NOTE
To prevent the accidental erasure of the client patient monitor’s
wireless LAN defaults, make sure that the server and client patient
monitors’ wireless LAN defaults match before transferring patient
monitor defaults.
1. At the server patient monitor, set up the monitor defaults you want
to store. Refer to the following sections of the operator’s manual for
reference:
Defaults Worksheet located in the preface of the manual. These
are blank and allow you to enter your own default settings.
Monitor Setup chapter.
Adult ICU, Neonatal ICU, Operating Mode Factory Default
Appendices.
2. When defaults are set, start from the Main Menu and select MORE
MENUS > MONITOR SETUP > SERVICE MODE.
3. Enter password using the Trim Knob control to select the day and
month from patient monitor screen with leading zeros. (e.g. July 4 =
0407)
4. Select MONITOR SETTINGS.
5. Select STORE DEFAULTS FOR NETWORK TRANSFER.
6. Select YES to the confirmation popup menu.

Table of Contents

Other manuals for GE Dash 3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 3000 and is the answer not in the manual?

GE Dash 3000 Specifications

General IconGeneral
BrandGE
ModelDash 3000
CategoryMonitor
LanguageEnglish

Related product manuals