Chapter 1. Calibration
12 DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
1.7.3 Entering Pulse Number
5. Enter the number of pulses desired (between 1 and 10,000) and press the [ENT] key. That number of pulses will
then be output at the specified frequency.
6. Repeat steps 3, 4 and 5 to test all four of the frequency/totalizer outputs. If any of the outputs fails to pass the test,
contact the factory for assistance.
Procedure Options
You have completed testing the totalizer/frequency outputs. Do one of the following:
• To calibrate/test additional slots, press [F4] (EXIT) and proceed to the appropriate section.
• To leave the Calibration Menu, press [EXIT] twice.