DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual 1
Chapter 1. Calibration
Chapter 1. Calibration
1.1 Introduction
Calibrating the Model DF868’s analog outputs and inputs is explained in this chapter. In addition, testing the optional
totalizer/frequency and alarm relay outputs is discussed. The following specific topics are included:
calibrating the built-in Slot 0 analog outputs
calibrating optional Slot 1 through Slot 6 analog outputs
calibrating optional Slot 1 through Slot 6 analog inputs
testing optional Slot 1 through Slot 6 alarm relays
testing optional Slot 1 through Slot 6 totalizer/frequency outputs.
The Model DF868 electronics console includes six expansion slots for the installation of option cards. These slots are
numbered 1-6, from right to left. In addition, every Model DF868 flowmeter includes two built-in analog outputs (A
and B) at terminal block
I/O, which is designated as Slot 0.
Note: Convention in this manual identifies any expansion slot as Slot x, where x is a number from 0-6.
See Chapter 1, Installation, of the Startup Guide for a complete description of the available option cards and the
procedures for wiring them.
1.2 Menu Map
The Calibration Menu is accessed by pressing the [CAL] key on the keypad. Use this menu to calibrate and test the Slot
0 analog outputs, as well as to calibrate and test any option cards that are installed in the expansion slots. Refer to the
menu map in Figure 6 on page 13 as a guide in following the calibration instructions.
Note: The same instructions apply to both the 1-Channel and 2-Channel versions of the Model DF868 flowmeter.
The following discussion assumes that the left screen pane is active. If the right screen pane is active, only the function
key designations change. That is, replace [F1]-[F4] with [F5]-[F8]. Proceed to the appropriate sections of this chapter to
calibrate and test all of the installed inputs and/or outputs.
Note: While in the Calibration Menu, if there is no keypad activity for two minutes, the Model DF868 will
automatically reboot and return to measurement mode.

Table of Contents

Other manuals for GE DigitalFlow DF868

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE DigitalFlow DF868 and is the answer not in the manual?

GE DigitalFlow DF868 Specifications

General IconGeneral
BrandGE
ModelDigitalFlow DF868
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals