Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 8 Replacement Procedures 8-23
Section 8-4Checks after FRU replacement (Debrief Guidelines)
Perform required Functional tests based upon the FRU being replaced.
Table 8-2 Functional Tests Matrix of LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
Section
FRU
No. Description Functional Tests
Basic Functional
Checks
4-3-1, 4-3-2, 4-3-7, 4-3-8, 4-3-9, 4-3-10
8-2-2 313 Handle Assy
Handle function check:
1. The system can be carried with the handle;
2. The handle can rotate from +90 to -110 degree
Section
8-3
Loading Base Image
Software
4-3-1, 4-3-2, 4-3-7, 4-3-8, 4-3-9, 4-3-10, 4-4, 8-3-3,
8-3-4, 8-3-5, 8-3-6, 8-3-7
Table 8-3 Functional Test key of LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
Key Functional Test/Diagnostics/Leakage Current
4-3-1 Power On/Boot Up
4-3-2
Power Off/ Shutdown
4-3-7 B Mode Checks
4-3-8 M Mode Controls
4-3-9 Color Flow Mode Checks
4-3-10
Doppler Mode Checks
4-4 Software Configuration Checks
8-3-3
Reload the Correct Preset Region
8-3-4
Option Strings Check
8-3-5
Probe Recognition Check
8-3-6
Peripheral Device Check
8-3-7
Reinstall DICOM Devices