Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 4 Functional Checks 4-13
4-3-8 M Mode Controls
4-3-8-1 Preparations
1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-28, in Chapter 3 Installation
to the System probe connector.
2.) Turn ON the scanner (if it isn’t turned on already).
Figure 4-9 Controls available in M Mode
Figure 4-10 M Mode Screen Picture Example
Primary Menu Secondary Menu
M Mode
Gain
M/D Cursor