7 Delivery checklist
1 Connecting cable
1 Hexagonal pin-type spanner
1 Operating instructions
1 Notes on moving the equipment safely
The rotor(s) and associated accessories are included in the delivery in the quantity ordered.
8 Unpacking the centrifuge
Lift the carton upward and remove the padding.
Do not lift by the front panel.
Observe the weight of the centrifuge, refer to chapter "Technical specifications".
Lift the centrifuge on both sides with an appropriate number of helpers and place it on the laboratory table.
9 Initial operation
A type B residual current circuit breaker must be used if the device is additionally protected with a residual current
circuit breaker in the building installation. When using a different type, the residual current circuit breaker may
either not switch off the unit if there is a fault on the unit, or it may switch off the unit even though there is no fault
on the unit.
Remove the transportation safety device from the bottom of the housing, see sheet ”Transportation safety device"
Position the centrifuge in a stable and level manner in a suitable place. During set-up, the required safety
margin of 300 mm around the centrifuge is to be kept according to EN / IEC 61010-2-020.
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous
substances or objects may be within the safety margin of 300 mm around the centrifuge.
Ventilation openings may not be blocked.
A distance of 300 mm must be maintained from the ventilation slots and openings of the centrifuge.
Check whether the mains voltage tallies with the statement on the type plate.
Connect the centrifuge with the power cord to a standard mains socket. For connection ratings refer to Chapter
"Technical specification".
Turn on the mains switch. Switch position "".
The machine type and program version will be displayed and the LEDs light up. After 8 seconds,
OPEN OEFFNEN is displayed and the left LED on the
STOP / OPEN
key blinks.
Open the lid.
The last used centrifuge data will be displayed.

Table of Contents

Other manuals for Hettich MIKRO 220R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich MIKRO 220R and is the answer not in the manual?

Hettich MIKRO 220R Specifications

General IconGeneral
BrandHettich
ModelMIKRO 220R
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals