EN
AB2200DEENFRIT Rev. 12 / 12.2022 47/107
16.3 Write protection for the programs
The programs can be protected against unintentional modification.
The write protection can be activated or deactivated as follows when the rotor is at a standstill:
Press and hold the key
SELECT
for 8 seconds.
After 8 seconds, SOUND / BELL appears in the display.
Press the
SELECT
key. The LOCK parameter is displayed.
Use the adjusting knob to set OFF or ON.
OFF: The programs are not write-protected.
ON: All programs are write-protected.
Write-protecting individual programs is not possible.
Press the
START / IMPULS
key to save the setting.
If ON is set,  lock  is displayed briefly as confirmation.
If OFF is set,  ok is displayed briefly as confirmation.
17 Centrifugation
When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous substances or
objects may be within the safety margin of 300 mm around the centrifuge.
If the permissible weight difference within the rotor loading has been exceeded, the drive shuts down during
the start-up, the unbalance display lights up, and IMBALANCE is displayed.
A centrifugation run can be stopped at any time by pushing the key
OPEN / STOP
.
All parameters can be selected and altered during the centrifugation run (see Chapter “Entering centrifugation
parameter”).
You can switch-over at any time between the RPM and RCF display with the key
RCF
. The input of the
centrifugation radius is necessary if you are working with the RCF display.
If OPEN OEFFNEN is displayed, a further operation of the centrifuge is only possible after opening the lid
once.
If R xx n-max xxxxx is displayed, then no centrifugation run has taken place as the rotor was changed
beforehand, refer to Chapter " Rotor Identification ".
Turn on the mains switch. Switch position .
Load the rotor and close the centrifuge lid.
17.1 Centrifugation with pre-set time
Adjusting time or recall a programme with pre-set time (see Chapter "Programming").
Press the key
START / IMPULS
. The LED in the button
START / IMPULS
lights up for as long as the rotor turns.
After expiration of the time or with truncation of the centrifugation run by pushing the key
OPEN / STOP
, the run-
down is effected with the selected brake step. The brake step is displayed.
During the centrifugation run, the rotor speed or the RCF value, the temperature in the centrifuging chamber (only for
centrifuge with cooling), and the remaining time are displayed.
17.2 Continuous run
Adjusting the symbol or recall a continuous run programme (see Chapter "Programming").
Press the key
START / IMPULS
. The LED in the button
START / IMPULS
lights up for as long as the rotor turns. The
time metering begins at 00:00.
Press the key
OPEN / STOP
in order to stop the centrifugation run. The run-down is effected with the selected
brake step. The brake step is displayed.
During the centrifugation run, the rotor speed or the RCF value, the temperature in the centrifuging chamber (only for
centrifuge with cooling), and the elapsed time are displayed.

Table of Contents

Other manuals for Hettich MIKRO 220R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich MIKRO 220R and is the answer not in the manual?

Hettich MIKRO 220R Specifications

General IconGeneral
BrandHettich
ModelMIKRO 220R
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals