3.1 Measurement Specifications
55
Specifications Chapter 3
3.1 Measurement Specifications
Common measurement specifications
Guaranteed
accuracy period
1 year (Opening and Closing of the Sensor: Maximum 10000
times)
Accuracy guaran-
tee for tempera-
ture and humidity
23°C±5°C (73°F±9°F) 80%RH or less (non-condensation)
Temperature
characteristics
-10 to 50°C Measurement accuracy x 0.1/°C (except 23°C±5°C)
Maximum rated
voltage to earth
600 VAC measurement category IV (anticipated tran-
sient overvoltage 8000 V)
Resistance measurement specifications
Guaranteed
accuracy
conditions
No reactance component, no noise current
Measurement
Method
Analog synchronous detection method (effective resistance mea-
surement)
Injected signal
frequency
Approx. 2.4 kHz
Injected voltage
level
Approx. 9.0 mV (with load open)
Effective
measuring range
0.02 to 1600
Zero suppression Less than 0.02
Overrange Greater than 1600
Measurement
response time
Filter: OFF/ON Approx. 3 sec /Approx. 9 sec.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents