SVC 4-61
Clean/Replace FPD Filter
` HAZARDOUS VOLTAGES ARE PRESENT IN THE INSTRUMENT WHEN THE
POWER CORD IS CONNECTED. AVOID A POTENTIALLY DANGEROUS SHOCK
HAZARD BY DISCONNECTING THE POWER CORD BEFORE WORKING ON THE
INSTRUMENT.
` FLAME PHOTOMETRIC (FPD) DETECTORS USE HYDROGEN GAS AS FUEL. BE
SURE ALL HYDROGEN GAS IS TURNED OFF AT ITS SOURCE BEFORE
REPLACING ANY FLOW COMPONENTS.
1. Set the main power line switch to the off
position.
2. Disconnect the power cable from its receptacle.
3. Turn off the hydrogen, air (or oxygen), and aux-
iliary gas supply to the detector by means of
the manifold on/off valves.
4. Allow time for the detector module to cool.
5. Remove the thumb-screw holding the detector
cover to the top of the HP 5890.
6. Facing the instrument, rotate the detector cover
to your right until the tab on the bottom left
edge of the cover comes free.
7. Raise the back of the cover and slide it towards
the rear of the instrument.
OFF ON
RED O VISIBLE
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 5890 Series II and is the answer not in the manual?

HP 5890 Series II Specifications

General IconGeneral
BrandHP
Model5890 Series II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals