SVC 4-1
Section 4
DETECTORS
REPLACING DETECTOR COMPONENTS
There are five detector options available for the HP 5890 Series II Gas Chromatograph; Thermal
Conductivity (TCD), Flame Ionization (FID), Nitrogen-Phosphorus (NPD), Electron Capture (ECD), and
Flame Photometric (FPD). Maintenance procedures for all the detectors are given in the following
pages. Procedures are supplied to remove, replace, and/or clean various subassemblies, based on the
current maintenance philosophy, i. e., to allow replacement of the lowest level components applicable
for a particular detector.
Specific part numbers are not given in this section. For all replacement part numbers, refer to Section 2
of the IPB portion of this document (Detector Options).
All of the detectors are heated using a heater/sensor setup which consists of at least one heater car-
tridge and one sensor cartridge. Heating of the detector zones is not covered in this section. For infor-
mation on the zone heater/sensor systems, refer to Section 6 of the service portion of this document
(Zone Temperature).
This document is not meant to provide instruction for first time installation of the detector options dis-
cussed. Add-on sheets exist for just this purpose, and should be referenced when performing a first
time installation.
TABLE OF CONTENTS
REPLACING DETECTOR COMPONENTS SVC 4-1. . . . . . . . . . . . . .
DETECTOR TROUBLESHOOTING SVC 4-3. . . . . . . . . . . . . . . . . . . .
THERMAL CONDUCTIVITY DETECTOR (TCD) SVC 4-4. . . . . . . . .
Remove/Replace TCD Detector Weldment SVC 4-4. . . . . . . . . . . . . . .. .. . . . . . . ..
Remove/Replace TCD Solenoid Switching Valve SVC 4-8. . . . . . . . . . . . . . .. .. . .
FLAME IONIZATION DETECTOR (FID) SVC 4-11. . . . . . . . . . . . . . . .
Remove/Replace FID Ignitor SVC 4-11. . . . . . . . . . . . . . .. .. . . . . . . .. . . . . . . . . . . . .
Remove/Replace FID Diode Bridge Assembly SVC 4-12.. . . . . . . . . . . . . . . .. . . . . .
Remove/Replace FID Collector Body/Collector Assembly SVC 4-14. . . . . . .. .. . . .
Remove/Replace FID Jet SVC 4-18. . . . . . . . . . . . . . .. .. . . . . . . .. . . . . . . . . . . . . . . .
Remove/Replace FID Detector Weldment SVC 4-21. . . . . . . . . . . . . . .. .. . . . . . . ..
NITROGEN-PHOSPHORUS DETECTOR (NPD) SVC 4-28. . . . . . . .
Remove/Replace NPD Active Element Power Transformer (Toroid) SVC 4-28. . . . .
Remove/Replace NPD Collector SVC 4-33. . . . . . . . . . . . . . .. .. . . . . . . .. . . . . . . . .
Remove/Replace NPD Jet SVC 4-35. . . . . . . . . . . . . . .. .. . . . . . . .. . . . . . . . . . . . . . .
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 5890 Series II and is the answer not in the manual?

HP 5890 Series II Specifications

General IconGeneral
BrandHP
Model5890 Series II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals