SVC 7-6
LIQUID SAMPLE VALVES (LSVs)
Liquid sampling valves are designed for use with liquefied gases under pressure such as ethane, propane,
butane, LNG, etc. They are not intended for nonvolatile liquids (at room conditions) where a concealed leak
may allow an accumulation or pool of liquid to form that may present a significant fire hazard. All standard
liquidsample valveshave1/16inchfittingsandare classifiedby thesamplesizeof theinstalledrotor(0.2mL,
0.5mL, or 1mL capacity). The two types of valves available are standard or low pressure (1000 psig) and
high pressure (5000 psig), in the four port, single purpose liquid sampling valve. Whenever a liquid sample
valveis used,an adjustablerestrictor is employed onthe sampleoutlet line to maintaininternal sample pres-
sure and thereby keep a compressed gas liquefied.