SVC 6-18
Remove/Replace Detector Zone Heater and Sensor Cartridges
` Remove/Replace TCD Heater and Sensor Cartridges
PERFORM THE PROCEDURE FOR DISCONNECTING/CONNECTING HEATER AND
TEMPERATURE SENSOR WIRING ON MAIN PCB BEFORE PERFORMING THIS
PROCEDURE.
1. Allow time for the oven and heated zones to cool.
2. At the bottom of the detector, inside the column oven,
remove the column and hardware associated with the
detector (liner, column/liner nuts, ferrules, makeup gas
adapter,etc.).
3. Cap the detector base.
4. Lift the hinged top cover at its front edge,
exposing the detector area.
5. Remove the TCD detector cover by removing two screws:
one from each side of the detector cover.
6. Cap the TCD vent port on the top of the detector. (This is not be required on a series con-
nected TCD.)
7. Remove the preformed thermal insulation from around the detector to expose the two screws
securing the detector to the instrument mainframe.
8. On a series-connected TCD, disconnect the TCD to FID jumper tube from the TCD oven-re-
turn exhaust vent port.
SOLENOID
VALVE
DETECTOR
COVER
SCREW
VENT
PORT
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 5890 Series II and is the answer not in the manual?

HP 5890 Series II Specifications

General IconGeneral
BrandHP
Model5890 Series II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals