SVC 3-20
11. Disconnect the EFS tubing from their connec-
tion sites at the tubing ends. (Label the tubes
as to their respective flow channel ( A or B)
to ensure an identical connection to the re-
placement EFS module.)
12. Free the EFS module by removing two screws.
The module can then be removed from its com-
partment in the side of the instrument.
13. Remove the connector from the EFS. Notice
the locking tab at each end of the plug. Re-
lease the locking tabs and pull the plug straight
out from its receptacle.
Locking Tabs
P11
WHENDISCONNECTINGA PLUG,PULL ON THE PLUG NOT ONITSWIRES. PULLING
ON THE WIRES MAY CAUSE BREAKAGE.
14. Remove the EFS module from the instrument.
15. Install the replacement EFS module in its compartment.
16. Secure the EFS module to the instrument using two screws.
17. Connect the plumbing to the EFS module.
18. Restore the gas supply to the instrument module (both channels, A and B), and check for
leaks at all installed fittings.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 5890 Series II and is the answer not in the manual?

HP 5890 Series II Specifications

General IconGeneral
BrandHP
Model5890 Series II
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals