Model 4342A
Section V
Paragraphs 5-9 to 5-12 '
5-9. PERFORMANCE CHECKS.
a. Connect an
AC
Voltmeter to 4342A LO
and
GND
terminals as shown in Figure
S-10. The Performance Checks compare the
4342A instrument with its specifications.
These checks are used in incoming inspection,
periodic maintenance, and after a repair.
Before beginning the Performance Checks, do
mechanical and electrical meter zero adjust-
ments using the procedure in Figure 3-6.
5-11. FREQUENCY ACCURACY CHECK.
An electronic frequency counter is required
for this check.
5-1. Connect Digital Voltmeter to
AC
Voltmeter dc output terminals.
b. Set 4342A controls as follows:
FREQUENCY RANGE
....... 70k - 220k
FREQUENCY dial
................. 20
Q RANGE
........................ 30
Q LIMIT
........................
CW
L/C dial ................... 25(pF)
AC dial .........................
0
a.
Connect frequency counter to 4342A
rear panel FREQUENCY MONITOR connector.
b.
Set 4342A controls as follows:
C.
FREQUENCY RANGE
........
22k to 70k
FREQUENCY dial ................ 2.2
Other controls
....... any settings
Frequency counter reading should be
within 21,678kHz to 22,320kHz.
d.
Check frequency at each frequency set-
ting in accord with Table 5-3. Counter
readings should be within the toler-
ance limits given in Table 5-3.
5-12.
Q RANGE CHECK.
An
AC
Voltmeter and a Digital Voltmeter are
required for this check.
C.
Set AC Voltmeter range to 30mV and
Digital Voltmeter to 1V. Digital
Voltmeter reading should be between
920.6 and 977.4mV.
d.
Set Q RANGE and
AC
Voltmeter range in
accord with Table 5-4.
Digital Volt-
meter reading should be within the
tolerance limits given in Table 5-4.
,
Table 5-4. Q Range Check.
1
( Q Range ( Vgzr / Digit:ztyter (
I
1000 1 mV 873.0 - 927.0 mV
DIGITAL VOLTMETER
uzl @I
G -
d DC OUTPUT
(REAR PANEL)
I
Figure 5-2. AQ Range Check.
5-3