Model 4342A
lower frequency limits to 9.850kHz
and lO.lSOkHz, respectively.
Para. 5-11 d.
Use Table 5-8 for op-
tion 001 instead of Table 5-3.
b.
Para. 5-12 b, 5-13 b, and 5-15 b.
Change FREQUENCY RANGE and FREQUENCY
dial settings to 1OOk - 320k and 2.0,
respectively.
5-29. OPTION 001 CALIBRATION AND
Para. S-20 j. Change frequency range
ADJUSTMENT PROCEDURES.
to 10 - 32kHz (from 22 - 70kHz).
S-30.
To apply the Calibration and Adjust-
ment Procedures in paragraphs 5-16 and those
below to option 001 instruments, partially
make the following changes in standard pro-
cedures:
a.
Para. 5-19 b.
Change the FREQUENCY
RANGE and FREQUENCY dial settings
to 10k - 32k and 2.0, respectively.
Section V
Paragraphs 5-29 and S-30
b. Para. S-20 b. Change the FREQUENCY
RANGE and FREQUENCY dial settings
to 10k - 32k and 1.0, respectively.
Para. S-20 e. Change frequency toler-
ance limits to lO.OOOkHz +O.lSOkHz.
Para. 5-20 f.
Change FREQUENCY dial
setting to 3.2.
Para. S-20 g. Change frequency toler-
ance limits to 32.000kHzi0.480kHz.
Para. 5-20k. Use Table 5-9 instead of
Table 5-7.
C.
Para. 5-21 b. Change FREQUENCY RANGE
and FREQUENCY dial settings to 10k -
32k and 1.0,
respectively.
d.
Para. 5-23 g. Change RF Oscillator
frequency setting to 32MHz.
Table 5-9. Frequency Adjustment (Option 001).
Frequent y
Range
10k - 32k
Frequency
Dial Setting
1.0
3.2
L
1.0
1OOk - 320k 3.2
L
1.0
1M - 3.2M 3.2
L
1.0
10M - 32M 3.2
L
3.2
32k - 1OOk 10
Ll
3.2
320k -
1M 10
L
3.2
3.2M - 10M 10
L
Measured
Frequency
Adjustment
10.000 +0.150
kHz
32.000 &O. 480
kHz
25.173 *to. 251
kHz
AlAlLl
AlAlCl
NONE
1
100.00 &l. 50 kHz
320.00 i4.80 kHz
251. ā73 i2.51 kHz
AlAlL
AlAlC5
NONE
1000.0 &5.0 kHz
3200.0 k48.0 kHz
2517.3 k25.1 kHz
AlAlL
AlAlC9
NONE
10.000 +O. 200 MHz
32.000 iO.640 MHz
25.1ā73 10.251 MHz
32.000 kO.480 kHz
100.00 *II. 50 kHz
79.618 *to. 796 kHz
AlAlL
AlAlC13
NONE
AlAlL
AlAlC3
NONE
AlAlL
AlAlC7
NONE
I
3200.0 *48.0 kHz
10.000 +O. 150 MHz
7961.8 i79.6 kHz
AlAlL
AlAlCll
NONE
5-13