55
15.6 Measurement Parameters
NIBP
Measurement method Oscillometric
Parameter display
Systolic pressure, diastolic blood, mean pressure and
pulse
Range of measurement for
adult
systolic pressure 40-270mmHg (5.3-36kPa)
diastolic pressure 10-215mmHg (1.3-28.7kPa)
mean pressure 20-235mmHg (2.7-31.3kPa)
Range of measurement for
child
systolic pressure 40-200mmHg (5.3-26.7kPa)
diastolic pressure 10-150mmHg (1.3-20kPa)
mean pressure 20-165mmHg (2.7-22kPa)
Range of measurement for
neonate
systolic pressure 40-135mmHg (5.3-18kPa)
diastolic pressure 10-100mmHg (1.3-13.3kPa)
mean pressure 20-110mmHg (2.7-14.7kPa)
Resolution 1mmHg (0.1kPa)
The measurement range and
accuracy of static pressure
0 mmHg (0 kPa) to 300 mmHg (40.0 kPa) to ±3 mmHg
(±0.4 kPa)
Accuracy ±5mmHg
Automatic Intervals
2, 3, 4, 5, 10, 15, 20, 25, 30, 60 or CO measurement
intervals
Note: Systolic and diastolic blood pressure measurements determined with this
device are equivalent to those obtained by a trained observer using the cuff/
stethoscope auscultation method, within the limits prescribed by the American
National Standard, Electronic or automated sphygmomanometers.
SpO2
Range and accuracy (70% to 100%*)
Own Brand 0% to 100%
±2% (measured without motion in adult/child mode)
±3% (measured without motion in adult/child mode)
Nellcor Module 0% to 100%
±2% (measured without motion in adult/child mode)
±3% (measured without motion in neonate mode)
Resolution 1%
* Accuracy 0% to 69% - not specifi ed
Specifi cations