hipot.com 14
4. Setting an AC Hipot Test
START
PROGRAM A MEMORY: TYPE AC
When TYPE is selected. Press the
rotary knob to edit the Test Type.
The options will blink.
Turn the yellow rotary knob to
scroll to desired parameter to
edit, TYPE.
Turn the rotary knob to change
the Test Type: ACW, DCW or IR.
Press the rotary knob to select
the Test Type.
You will be returned to the
previous menu.
START
PROGRAM A MEMORY: CHANGE VOLTAGE
When VOLT is selected. Press the
rotary knob to edit the Voltage.
The rst digit of Voltage will blink.
Turn the yellow rotary knob to
scroll to desired parameter to
edit VOLT.
Turn the rotary knob to change
the Voltage setting for each digit.
Press the rotary knob to select the
setting and move to the next digit.
When done, you will be returned
to the previous menu.