6.3.2 Flash Codes
Number of
flashes
Fault Impact on
Scooter
Notes
1
Battery needs
charging
Will drive
Battery charge is running low. Recharge
the batteries as soon as possible.
2
Battery voltage
too low
Drive inhibited
Battery charge is empty. Recharge the
batteries.
If the scooter is left off for a few minutes,
battery charge may recover sufficiently to
allow driving for a short period of time.
Lifter 9raised
Driving speed
reduced
Lower lifter completely.
3
Battery voltage
too high
Drive inhibited
Battery charge is too high. If a charger is
plugged in, unplug it.
The electronic system charges the
batteries when travelling down slopes or
decelerating. Excessive charging in this
manner may cause this fault. Turn the
scooter power off and then back on
again.