Basic Operation
The IT6000C series power supply supports two connection methods: Local
measurement and Remote sensing. The remote sensing is used for maximizing
measurement accuracy. (Refer to 2.5 Connecting the Device Under Test (DUT))
The procedures to set the menu item are as follows.
1. Press the composite keys [Shift]+[P-set] (System) on the front panel to en-
ter the system menu.
2. Press the Up/Down key or turn the knob to select the Sense and press
[Enter].
3. Press the Left / Right key or turn the knob to adjust the value of this
parameter.
• Off: Default value, indicates turn the sense function off.
• On: Indicates turn the sense function on.
4. After the parameter settings are complete, press [Enter].
5.8 Select Trigger Source (Trig Source)
For the IT6000C series power supply, the List and data logging functions can be
triggered for running by the following methods:
• Immediate: Perform a trigger operation immediately.
• Manual: Default value, indicates the trigger occurs when the
[Shift]+[On/Off] (Trigger) keys are pressed from the front panel.
• Bus: Indicates the trigger occurs via the communication cables. When the
trigger command *TRG is received, the IT6000C instrument generates a
trigger.
• Voltage (only available in DLogTrig Source): Voltage trigger.
When the DC terminals detect that the voltage reaches the trigger voltage
setting value and is within the range of the upper and lower trigger limits, a
data recording operation is triggered.
– Level: Trigger threshold.
– Up-Level: Trigger upper limit value.
– Down-Level: Trigger lower limit value.
• Current (only available in DLogTrig Source): Current trigger.
When the DC terminals detect that the current reaches the trigger current
setting value and is within the range of the upper and lower trigger limits, a
data recording operation is triggered.