IEEE488.2 Commands
Copyright © ITECH Electronic Co., Ltd. 22
*RCL
This command restores the electronic load to a state that was previously stored
in memory with a *SAV command to the specified location. All states are
recalled with the following exceptions:
CAL: STATe is set to OFF
The trigger system is set to the Idle state by an implied ABORt command (this
cancels any uncompleted trigger actions)
NOTE: The device state stored in location 0 is automatically recalled at power
turn-on.
Command Syntax
*RCL <NRf>
Parameters
0 to 9
Examples
*RCL 3
Related Commands
*PSC *RST *SAV
*RST
This command reset the electronic load to the factory-defined states.
Command Syntax
*RST
Parameters None
*SAV
This command stores the present state of the electronic load to a specified
location in memory. Up to 100 states can be stored. If a particular state is
desired at power-on, it should be stored in location 0. It then will be recalled at
power-on if the power-on state is set to RCL0. Use *RCL to retrieve instrument
states.
NOTE: *SAV does not save the programmed trigger values
([SOURce:]CURRent:TRIGGer, [SOURce:]RESistance:TRIGGer,
[SOURce:]VOLTage:TRIGGer). Programming an *RCL or a *RST command
causes the triggered settings to revert to their [IMMediate] settings.
Command Syntax
*SAV <NRf>
Parameters
0 to 100
Examples
*SAV 3
Related Commands
*PSC *RST *RCL
*SRE
This command sets the condition of the Service Request Enable Register. This
register determines which bits from the Status Byte Register (see *STB for its bit
configuration) are allowed to set the Master Status Summary (MSS) bit and the
Request for Service (RQS) summary bit. A 1 in any Service Request Enable
Register bit position enables the corresponding Status Byte Register bit and all
such enabled bits then are logically OR to cause Bit 6 of the Status Byte
Register to be set.

Table of Contents

Other manuals for ITech IT8811

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8811 and is the answer not in the manual?

ITech IT8811 Specifications

General IconGeneral
BrandITech
ModelIT8811
CategoryTest Equipment
LanguageEnglish

Related product manuals