Measure Commands
Copyright © ITECH Electronic Co., Ltd. 34
MEASure:VOLTage[:DC]?
This command initiates and executes a new voltage measurement, and returns the
measured input voltage of the electronic load.
Group
Measurement
Syntax
MEASure:VOLTage[:DC]?
Returns
<NR2> is the measured input voltage in volts.
Example
MEAS:VOLT? might return 5.0011, which would be the
measured voltage across the electronic load inputs in volts.
MEASure:VOLTage:MAX?
This command initiates and executes a new maximum voltage measurement, and
returns the measured input maximum voltage of the electronic load.
Group
Measurement
Syntax
MEASure:VOLTage:MAX?
Returns
<NR2> is the measured input maximum voltage in volts.
Example
MEAS:VOLT:MAX? might return 100.0011, which
would be the measured maximum voltage across the
electronic load inputs in volts.
MEASure:VOLTage:MIN?
This command initiates and executes a new minimum voltage measurement, and
returns the measured input minimum voltage of the electronic load.
Group
Measurement
Syntax
MEASure:VOLTage:MIN?
Returns
<NR2> is the measured input minimum voltage in volts.
Example
MEAS:VOLT:MIN? might return 1.0011, which would
be the measured minimum voltage across the electronic
load inputs in volts.
MEASure:CURRent[:DC]?
This command initiates and executes a new current measurement, and returns the
measured input current of the electronic load.

Table of Contents

Other manuals for ITech IT8811

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8811 and is the answer not in the manual?

ITech IT8811 Specifications

General IconGeneral
BrandITech
ModelIT8811
CategoryTest Equipment
LanguageEnglish

Related product manuals