Source Subsystem
Copyright © ITECH Electronic Co., Ltd. 47
Returned Parameters <NR1>
Related Commands CURR:PROT CURR:PROT:STAT
[SOURce:]CURRent:TRANsient:MODE
This command selects the operating mode of the transient generator as follows
in constant current mode.
CONTinuous The transient generator puts out a continuous pulse stream
after receipt a trigger signal.
PULSe The transient generator puts out a single pulse upon receipt of a
trigger.
TOGGle The transient generator toggles between two levels upon receipt of
a trigger.
Command Syntax [SOURce:]CURRent:TRANsient:MODE <mode>
Parameters CONTinuous | PULSe | TOGGle
*RST Value CONTinuous
Examples CURR:TRAN:MODE TOGG
Query Syntax [SOURce:]CURRent:TRANsient:MODE?
Returned Parameters <CRD>
Related Commands CURR:TRAN:ALEV TRAN
[SOURce:]CURRent:TRANsient:ALEVel
[SOURce:]CURRent:TRANsient:BLEVel
This command specifies the transient level of the input current. The transient
function switches between the level a and level b.
Command Syntax [SOURce:]CURRent:TRANsient:ALEVel <NRf+>
[SOURce:]CURRent:TRANsient:BLEVel <NRf+>
Parameters 0 through MAX | MINimum | MAXimum | DEFault
Unit A (amperes)
*RST Value ALEVEL MAXimum , BLEVel MINnum
Examples CURR:TRAN:ALEV 5 CURR:TRAN:BLEV 0.5
Query Syntax [SOURce:]CURRent:TRANsient:ALEVel?
[ MINimum | MAXimum | DEFault ]
[SOURce:]CURRent:TRANsient:BLEVel?
[ MINimum | MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands CURR
[SOURce:]CURRent:TRANsient:AWIDth
[SOURce:]CURRent:TRANsient:BWIDth
This command specifies the transient pulse width of the input current.
Command Syntax [SOURce:]CURRent:TRANsient:AWIDth <NRf+>
[SOURce:]CURRent:TRANsient:BWIDth <NRf+>
Parameters 20 to 3600 us
Unit S (second)
*RST Value 500uS