Source Subsystem
Copyright © ITECH Electronic Co., Ltd. 45
*RST Value MAXimum (high range)
Examples SOUR:CURR:RANGE MIN
Query Syntax [SOURce:]CURRent:RANGe? [ MINimum |
MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands CURR CURR:SLEW
[SOURce:]CURRent:SLEW
This command sets the slew rate for all programmed changes in the input
current level of the electronic load. This command programs both positive and
negative going slew rates. MAXimum sets the slew to the fastest possible rate.
MINimum sets the slew to the slowest rate.
Command Syntax [SOURce:]CURRent:SLEW[:BOTH] <NRf+>
Parameters MINimum to MAXimum | MAXimum | MINimum |
DEFault
Unit A (amps per micro second)
*RST Value MAXimum
Examples CURR:SLEW MAX
Related Commands CURR:SLEW:NEG CURR:SLEW:POS
[SOURce:]CURRent:SLEW:POSitive
This command sets the slew rate of the current for positive going transitions.
MAXimum sets the slew to the fastest possible rate. MINimum sets the slew to
the slowest rate.
Command Syntax [SOURce:]CURRent:SLEW:POSitive <NRf+>
Parameters MINimum to MAXimum | MAXimum | MINimum |
DEFault
Unit A (amps per micro second)
*RST Value MAXimum
Examples CURR:SLEW:POS MAX
Query Syntax [SOURce:]CURRent:SLEW:POSitive? [ MINimum |
MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands CURR:SLEW
[SOURce:]CURRent:SLEW:NEGative
This command sets the slew rate of the current for negative going transitions.
MAXimum sets the slew to the fastest possible rate. MINimum sets the slew to
the slowest rate.
Command Syntax [SOURce:]CURRent:SLEW:NEGative <NRf+>
Parameters MINimum to MAXimum | MAXimum | MINimum |
DEFault
Unit A (amps per micro second)
*RST Value MAXimum
Examples CURR:SLEW:NEG MAX
Query Syntax [SOURce:]CURRent:SLEW:NEGative? [ MINimum |
MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands CURR:SLEW