EasyManuals Logo
Home>LeCroy>Test Equipment>HDO6000

LeCroy HDO6000 User Manual

LeCroy HDO6000
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
HDO6000 High Definition Oscilloscope
Determining Differential Time Measurements
The instrument enables accurate differential time measurements between two traces: for example,
propagation, setup and hold delays (see figure).
If included in your oscilloscope, parameters such as Delta c2d± require the transition polarity of the clock
and data signals to be specified.
Moreover, a hysteresis range may be specified to ignore any spurious transition that does not exceed the
boundaries of the hysteresis interval. In the figure, Delta c2d- (1, 2) measures the time interval separating
the rising edge of the clock (trigger) from the first negative transition of the data signal. Similarly, Delta
c2d+ (1, 2) measures the time interval between the trigger and the next transition of the data signal.
68
922499 Rev B

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LeCroy HDO6000 and is the answer not in the manual?

LeCroy HDO6000 Specifications

General IconGeneral
BrandLeCroy
ModelHDO6000
CategoryTest Equipment
LanguageEnglish

Related product manuals