HDO6000 High Definition Oscilloscope
l On theTrack right-hand dialog, uncheck Auto Find Scale and enter a new Center and Height/div.
View Trend
This procedure explains how to view the trend of a measurement parameter. A trend is a waveform
composed of a series of parameter measurements in the order the measurements were taken. The
vertical units are those of the source parameter, the horizontal unit is measurement number. The trend
contains a single value for each measurement.
Although the trend plots measurement values, the plot is drawn as a math function and controlled through
the Math dialog.
1. If you're not already on the Measure dialog, choose Measure Measure Setup....
2. Touch the Px tab for the parameter you wish to plot.
3. Touch the Trend buttonat the bottom of the dialog and choose a math function Fx in which to draw the
Trend.
The Trend is displayed in a new grid, along with its function descriptor box.
4. To rescale the Trend plot:
l Touch the Trend function descriptor box to open the Math dialog, then touch the Trend tab at the
far right of the dialog.
l On the Trend right-hand dialog, uncheck Auto Find Scale and enter the new Center and Height
values.
86
922499 Rev B

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LeCroy HDO6000 and is the answer not in the manual?

LeCroy HDO6000 Specifications

General IconGeneral
BrandLeCroy
ModelHDO6000
CategoryTest Equipment
LanguageEnglish

Related product manuals