14.1.3 Wear
Check all bolts and tighten if necessary.
Check all locking mechanisms.
Check the bed for wear, scratches or rub marks.
Eliminate the cause if necessary.
Have any defective parts replaced.
14.2 Functioning
Check that all bed adjustments reach the maximum position.
If necessary, clean, lubricate or replace any worn areas and parts.
14.2.1 Electric Control
Plug Connections
Replace O-rings on connectors.
Check the plug connections for dirt and defects.
Clean or replace if necessary.
Check the plug connectors are properly seated.
Motors
Check motor movement (adjust bed positions). Check for incorrect and interrupted movements.
Have defective motors replaced if necessary.
Check cables for signs of wear and entanglement.
Install new cable or replace the motor if necessary.
Battery
Check the battery is working properly (disconnect the bed from the mains).
Replace the battery if necessary.
14.2.2 Castors
Clean the castors completely.
Grease the castors if necessary. Use Caro EP 2 by DEA or an equivalent grease
Check the castors work properly.
o Forward Movement
o Unrestricted Movement
o Braked
Have the brakes adjusted if necessary.
Have any defective castors replaced.
14.2.3 Accessories
Check all accessories (for example, lifting pole, side rails, infusion stand, etc.) are working properly.
Replace if necessary.