Service, maintenance and technical support
64 Snow depth sensor SHM 31, V2.1
Error codes
Description
E61
Hardware error in the interface
E62
Incorrect value in the interface communication (SIO parity error)
E63
SIO overflow; check time for output signals in application software
E64
SIO framing error; serial interface parameter not set correctly to 8N1
E65
Evaluation routine: In some cases, measurements in the calculation interval were
ignored because they would have exceeded the maximum permitted change in snow
depth.
E66
Evaluation routine: The most recently valid snow depth was output, as all
measurements in the calculation interval would have exceeded the maximum
permitted snow depth change.
E67
Measurement was cancelled by ‘MEN’
E68
No valid telegram available yet (e.g. after starting the measurement with ‘MST’)
E70
Evaluation routine could not read settings
E71
Evaluation routine has not received any data from the laser
E72
Evaluation routine has no valid laser temperature values
E73
Evaluation routine has no valid block temperature values
E74
Evaluation routine has no valid outside temperature values
E75
Evaluation routine has no valid laser distance measurement values
E76
Evaluation routine: G-sensor vector is an invalid length
E77
Evaluation routine is using the reference angle, as the current angle is invalid
E78
Evaluation routine: Signal calibration: signal_high <= signal_low
E79
Evaluation routine: Signal calibration: Signal too small
E80
Evaluation routine: Signal calibration: Signal too large
E81
Evaluation routine: Signal calibration: no angle correction, as angle > 90 degrees
E82
Evaluation routine: channel_average_count too large
E83
Evaluation routine could not initialise ring buffer for avg / min / max channels
Table 27: Error codes
13.4. UMB status codes
The UMB status codes are also used for error and status analysis. A description can be found in
the general UMB description.

Table of Contents

Other manuals for Lufft SHM31

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lufft SHM31 and is the answer not in the manual?

Lufft SHM31 Specifications

General IconGeneral
BrandLufft
ModelSHM31
CategoryAccessories
LanguageEnglish

Related product manuals