9-8 BeneLink Module Operators Manual
9.3.4 Setting the Serial Ports of the Pumps
Brand Model Protocol Baud Rate Data Bits Stop Bits Parity Remark
B.Braun Perfusor Space / 57600 8 1 None Using the Type F cable to connect the
spaceCom of the spaceStation (if
cascading is needed. Refer to the
official documents on how to cascade
several spaceStation. Use an PC to
connect the docking station through
Ethernet Open IE and input
192.168.100.41 Configuration
BCC configuration panel : BCC 3.3x ,
interface: com1, Baudrate:57600, Stop
bits: 1, Databits: 8, Character Stuffing:
On, Operation Mode: Request
Fresenius
Fresenius Pumps
/ 115200 8 1 None No need to set