System Maintenance 16-5
CAUTION:
Please clean all dust-proof covers of the system periodically
(1 time per month); otherwise, system damage may result.
Cleaning times can be increased when the system is used in
the open air or somewhere dust is more.
16.1.2 Clean the Peripherals
Do the cleaning maintenance according to your actual peripheral configuration; items
which are not configured can be skipped.
Content Description
Color and B/W
video printer
First wipe off dust or stain attached to the cover of printer with soft dry
cloth, then clean the inside of printer. Be sure to do the cleaning
maintenance according to the operation manual if is necessary.
Graph / text
printer
First wipe off dust or stain attached to the cover of printer with soft dry
cloth, then clean the inside of printer. Be sure to do the cleaning
maintenance according to the operation manual if is necessary.
Footswitch
Use soft dry cloth with a little mild soap water to wipe off the dust or
stain attached to the pedals or cable of foot switch.
Bar code
scanner
First use soft dry cloth to wipe off dust attached to glass panel of
scanner, then the dust or strain attached to cable and bracket. Be sure
to do the especial cleaning maintenance according to the operation
manual if is necessary.
16.1.3 Checking the Probe
îš„ Visually check to confirm that there is no crack and expansion to probe head.
îš„ Visually check to confirm that there is no deterioration or desquamation to probe
cable.
îš„ Visually check to confirm that there is no bend, destroyed or falling off pins to the
connector.
16.1.4 Checking the Power Cable and Plug
îš„ Visually check to confirm that there is no wrinkles, crack or deterioration; No crack
and expansion to the surface of adapter.
îš„ Manually check to confirm that there is no looseness or rupture. The connection of
plug is reliable.
16.1.5 Checking Appearance
Check if there is any crack in the covers: