3.1.1 Test Report
After completing the tests, service personnel are required to record test results in the
following table and report them to our Customer Service Department.
Test Equipment
Name Model/PN Expiry Date
Test Record
No. Test item Test point Test Results
1
2
Conclusion
Pass/Fail: Tested by: Date:
3.1.2 Recommended Frequency
Check/Maintenance Item Frequency
Visual test When first used or not used for a prolonged time.
Power on test 1. When first used or not used for a prolonged time.
2. Following each repair or replacement of main unit
part.
SpO
2
test
PR test
1. When you suspect that the measurement is inaccurate.
2. Following each repair or replacement of the SpO
2
module.
3. At least once every two years.
Real-time data exporting test
Trend data exporting test
When you suspect that data exporting does not work
properly.
Enclosure leakage
current test
Electrical safety
tests
Patient leakage
current test
At least once every two years.