EasyManuals Logo
Home>Mindray>Medical Equipment>Trio

Mindray Trio Operating Instructions

Mindray Trio
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #129 background imageLoading...
Page #129 background image
Temperature Sensor Cleaning and Disinfection (Reusable) User Maintenance
4 - 6 0070-10-0666-01 Trio™ Operating Instructions
4.5 Temperature Sensor Cleaning and Disinfection
(Reusable)
The temperature probe should not be heated above 100 ºC (212 ºF). It should only be
subjected briefly to temperatures between 80 ºC (176 ºF) and 100 ºC (212 ºF)
The probe must not be sterilized in steam
Only detergents containing no alcohol can be used for disinfectant
The rectal probes should be used, if possible, in conjunction with a protective rubber
cover
To clean the probe, hold the tip with one hand. With the other hand rub the probe down
in the direction of the connector using a moist lint-free cloth
NOTE: Disposable temperature probe must not be re-sterilized or
reused.
4.6 Battery Replacement and Maintenance
Battery Replacement
1. Open the battery compartment door located on the left side panel of the monitor by
using the finger grip area and sliding the door backward.
2. Lift the release latch in the upper right corner of the battery compartment. This will
release the battery. Remove the battery.
3. Insert the replacement battery, contact end first, with the positive (+) contact facing
downward. Slide in the replacement battery until it latches.
4. Close battery compartment door by sliding the door forward until it firmly locks into
place.
CAUTION: Replace the battery with one of the following part numbers:
0146-00-0043 (for a sealed lead acid battery), 0146-00-
0069 (for a Lithium Ion battery).
Battery Maintenance
The batteries used in the Trio monitor are of sealed lead acid construction and Lithium Ion
construction. These types of batteries may be subject to local regulations regarding disposal.
At the end of battery life, dispose of the batteries in accordance with any local regulations.
CAUTION: Remove the battery if the Trio is not likely to be used for an
extended period of time.
CAUTION: Remove the battery prior to shipping the Trio.

Table of Contents

Other manuals for Mindray Trio

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Trio and is the answer not in the manual?

Mindray Trio Specifications

General IconGeneral
BrandMindray
ModelTrio
CategoryMedical Equipment
LanguageEnglish

Related product manuals