EasyManuals Logo
Home>Mindray>Medical Equipment>Trio

Mindray Trio Operating Instructions

Mindray Trio
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
Trio™ Operating Instructions 0070-10-0666-01 6 - 7
Appendix Patient Parameter Specifications
6.2.3 ANSI/AAMI EC13-2002 Compliance
The ECG meets the section 4.2 Performance Requirements of ANSI/AAMI EC13-2002
except for paragraph 4.2.9.13, Synchronizing Pulse for Cardioversion, which is not
applicable.
6.2.3.1 Disclosed Performance Specifications
Tall T-Wave Rejection: When tested in accordance with ANSI/
AAMI EC13-2002 section 4.1.2.1 c), the
heart rate meter will reject all T-waves with
amplitudes less than 120% of a 1 mV, 100
ms QRS, and a T wave duration of 180 ms
and a Q-T interval of 350 ms.
Heart Rate Averaging: ANSI/AAMI EC13-2002 section 4.1.2.1 d).
The Heart Rate Averaging computation is as
follows: The average of the last 4 R-to-R
intervals, when last 3 R-to-R intervals > 1200
ms. Otherwise, the average of the last 12 R-
to-R intervals, minus the maximum and
minimum values. The update rate of the
Heart Rate on the display is once per
second.
Heart Rate Meter Accuracy and Response to
Irregular Rhythm:
When tested in accordance with ANSI/
AAMI EC13-2002 section 4.1.2.1 e), the
indicated heart rate after a 20 second
stabilization period is:
Figure 3a (Ventricular Bigeminy) - 80 bpm
Figure 3b (Slow Alternating Ventricular
Bigeminy) - 60 bpm
Figure 3c (Rapid Alternating Ventricular
Bigeminy) - 120 bpm
Figure 3d (Bi-directional Systoles) - 90 bpm

Table of Contents

Other manuals for Mindray Trio

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Trio and is the answer not in the manual?

Mindray Trio Specifications

General IconGeneral
BrandMindray
ModelTrio
CategoryMedical Equipment
LanguageEnglish

Related product manuals