Chapter 4 Analog Input
X Series User Manual 4-52 ni.com
common-mode rejection ratio (CMRR). These methods are outlined in the
Connecting Analog Input Signals section.
AI GND is an AI common signal that rou tes directly to the ground
connection point on the devices. You can use this signal if you need a
general analog ground connection point to the device.
Analog Input Timing Signals
In order to provide all of the timing functionality described throughout this
section, Simultaneous MIO X Series devices have a flexible timing engine.
Refer to the Clock Routing section of Chapter 9, Digital Routing and Clock
Generation.
Simultaneous MIO X Series devices use AI Sample Clock
(ai/SampleClock) to perform simultaneous sampling on all active analog
channels. Since there is one ADC per channel, AI Sample Clock controls
the sample period on all the channels in the task.
An acquisition with posttrigger data allows you to view data that is acquired
after a trigger event is received. A typical posttrigger DAQ sequence is
shown in Figure 4-28. The sample counter is loaded with the specified
number of posttrigger samples, in this example, five. The value decrements
with each pulse on AI Sample Clock, until the value reaches zero and all
desired samples have been acquired.
Figure 4-28. Typical Posttriggered DAQ Sequence
An acquisition with pretrigger data allows you to view data that is acquired
before the trigger of interest, in addition to data acquired after the trigger.
Figure 4-29 shows a typical pretrigger DAQ sequence. The AI Start Trigger
signal (ai/StartTrigger) can be either a hardware or software signal. If
AI Start Trigger is set up to be a software start trigger, an output pulse
appears on the ai/StartTrigger line when the acquisition begins. When the
AI Start Trigger pulse occurs, the sample counter is loaded with the number
of pretrigger samples, in this example, four. The value decrements with
each pulse on AI Sample Clock, until the value reaches zero. The sample
13042
AI Start Trigger
AI Sample Clock
Sample Counter
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments NI 6323 and is the answer not in the manual?

National Instruments NI 6323 Specifications

General IconGeneral
BrandNational Instruments
ModelNI 6323
CategoryI/O Systems
LanguageEnglish

Related product manuals