Instructions for Use XLTEK Protektor32
34
Environmental Conditions
Operating Environmental
Limits
Temperature: 10° to 40°C
Relative Humidity: 30%75%
Altitude: Up to 4600 meters above sea level
Atmospheric Pressure: 500 hPa to 1060 hPa
Transport and Storage Limits Temperature Range: -40°C to 70°C
Humidity Range: 10%100%, including condensation
Atmospheric Pressure Range: 500 hPa to 1,060 hPa
Electrical Ratings of Isolation
Box, Acquisition Box, and
Stimulator Box with
Protektor32 Cart
24Vdc/ 2A
Isolation Transformer
Maximum Output Power
P/N 006979: 100VAC-50/60Hz, 8.7/4.35A (850VA),
115(120)/230(240)VAC-50/60Hz, 8.7/4.35A (1000VA)
P/N 010758: 120VAC-60Hz, 7.50A (900VA)
Waste Management
Waste Electrical and Electronic Equipment Compliance.
Do not dispose of this product in the unsorted municipal waste stream. Dispose of the
product according to the local regulations.