32
WB135745-W10
12 Electromagnetic compatibility (EMC)
12.1 Guidance and manufacturer’s declaration –
electromagnetic emissions
The peristaltic pump unit is intended for use in the environment specified bellow. The
user of the peristaltic pump unit should assure that it is used in such an environment.
Emission test Compliance
Electromagnetic environment –
guidance
High frequency emissions
CISPR 11
Group 1 The peristaltic pump unit uses high
frequency signals only for its
internal function. Therefore, its
high frequency emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
High frequency emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations / flicker
emissions IEC 61000-3-3
complies
The peristaltic pump unit is suitable
for the use in all establishments,
including domestic establishments
and those directly connected to the
public-low-voltage power supply
network that supplies buildings
used for domestic purposes.
Length of power connection line: ≤ 3 m
CAUTION
The peristaltic pump unit should not be used adjacent to or stacked with
other electrical equipment, except devices which are intended for this
purpose and tested by the manufacturer.
The use of accessories which are not approved by the manufacturer
may result in an increase of electromagnetic emissions and the
compliance with the stipulated limit values are not guaranteed anymore.