Instructions for Use EN Olympus AFU-100
34
WB135745-W10
Immunity test IEC 60601-1-2
test level
Compliance
level
Electromagnetic
environment – guidance
Portable and mobile high
frequency communication
equipment should be used not
closer to any part of the
peristaltic pump unit, including
cables, then the recommended
separation distance calculated
from the equation applicable to
the frequency of the
transmitter; where P is the
maximum output power rating
of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m):
Conducted high
frequency
according to
IEC 61000-4-6
3 V
eff
150 kHz to
80 MHz
3 V 1.2 * Square root (P)
Radiated high
frequency
according to
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m 1.2 * Square root (P)
for 80 … 800 MHz
2.3 * Square root (P)
for 800 MHz … 2.5 GHz
The field strength from fixed
high frequency transmitters, as
determined by survey, should
be less than the compliance
level.
Interference may occur in the vicinity of equipment marked with the symbol “non-
ionising radiation“.
NOTE
This guideline may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.