USG-400
Safety check 6-6 ISSUE1
2 Check Card
Date of check
Product
SN
Inspected by
Checked by
Inspection item Criteria
Measurement
Equipment No.
result
Grounding Resistance The resistance must be 0.1
Normal Condition
500 μA
μA
Ground leakage
current
Single Fault
1000 μA
μA
Normal Condition
100 μA
μA
Enclosure leakage
current
Single Fault
500 μA
300 μA (UL Deviation)
μA
AC 10 μA
μA
Normal Condition
DC 10 μA
μA
AC 50 μA
μA
Patient leakage
current
Single Fault
DC 50 μA
μA