10.4. Printout of all balance parameters through RS 232 port
After pressing F ke
y, balance parameters will be send through RS 232 port.
Return to weighing
(see - 7.1.2. – return to weighing).
11. BALANCE CALIBRATION
As the
acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated – in compliance with the underlying physical
weighing principle – to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory).
This adjustment process must be carried out for the location in the factory). This
adjustment process must be carried out for the commissioning, after each change
of location as well as in case of fluctuating environment temperature. It is also
recommend to adjust the balance periodically during weighing operation in order to
obtain exact measured values.
To ensure high precision of weighing corrective factor in relation to standard mass
must be noted in the balance memory periodically – it is the balance calibration.
Calibration should be performed:
- Before the weighing,
- When long breaks are between following measure series
- When temperature inside the balance changes more than: 3° C.
Kind of calibration:
- Internal automatic calibration
* started if temperature changes
* started if the time changes

Table of Contents

Other manuals for RADWAG PS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG PS Series and is the answer not in the manual?

RADWAG PS Series Specifications

General IconGeneral
BrandRADWAG
ModelPS Series
CategoryScales
LanguageEnglish

Related product manuals