15.2. Printing data with date and time
Every printout of weighing can be printed with date and time of measurement.
It’s possible if balance is connected to printer Kafka 1/Z or Kafka SQ S.
After connecting balance and printer set in balance menu parameters of P2
GLP group:
PdAt on value YES
Ptin on value YES
15.3. Cooperation with statistic printout Kafka SQS
After conn
ecting balance to printer KAFKA SQ S statistic of measurement is
possible to do. Example of printout with statistics from series of measurement:
1 9:02:15 + 7.0016 g
2 9:02:39 + 5.0152 g
3 9:02:58 +
12.0171 g
4 9:03:15 + 9.9937 g
5 9:03:34 + 12.0169 g
6 9:03:48 + 22.0111 g
Data 13.09.2001 Godz. 9:04
n 6 batch quantity
sum x 68.0556 g summary of mass of samples
x 11.34260 g average value
s 5.92328 g standard deviation
srel 52.22 % factor of variancy
min 5.0152 g min value
max 22.0111 g max value
R 16.9959 g difference max – min
15.4. Format of sending data
Result of
weighing can be sent from balance to external device after pressing
the PRINT button on the balance or after sending order from the computer.

Table of Contents

Other manuals for RADWAG PS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG PS Series and is the answer not in the manual?

RADWAG PS Series Specifications

General IconGeneral
BrandRADWAG
ModelPS Series
CategoryScales
LanguageEnglish

Related product manuals