18. ERROR MESSAGES
Er1 Hi
- wrong initial mass or calibration
Er2 nuLL - values from A.D converter out of the normal range - under range
Er3 FuL1 - values from A.D converter out of the normal range - over range
Er4 FuL2 - over range
Er5 rout - value out of range for performed function e.g value > +/- 2% from initial
mass for zeroing,
e.g. value <= 0 for tarring
Er7 tout - timeout error for zeroing and tarring
Er8 outr - enter value out of range
Er9 Lock - function locked e.g. to protect function if LFt (legal for trade) is on,
Er10 cal - calibration error (change of mass or incorrect mass of standard mass).
19. ADDITIONAL EQUIPMENT
19.1. Vibration damping table
T
hanks to this stable base all vibrations are eliminated.Inside of
table marble plate is installed which makes foundation for balance.
19.2. Rack for weighing load under balance
It i
s used when weighing under balance is neccesary. It is usefull
for magnetic load or when thickness of products is defined.
Racks are used also during tests for absorptiveness of materials
e.g. absorptiveness of foamed polystyrene. Structre of rack is
made of steel. Rack is 330mm high.
19.3. Set for defining thickness of solids and liquids
It’s appropr
iate for weighing with precision 1mg. It enables to
define thickness solids and liquids. This procedure is fully
automatically user only puts samples on weighing pans.

Table of Contents

Other manuals for RADWAG PS Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG PS Series and is the answer not in the manual?

RADWAG PS Series Specifications

General IconGeneral
BrandRADWAG
ModelPS Series
CategoryScales
LanguageEnglish

Related product manuals