4.3. Leveling
Before plugging to mains, level the balance by turning the
adjustable feet in a way that the air bubble of the level is in its
central ring. The level is located at the back wall of balance’s
housing.
4.4. Standard equipment of a balance
• Balance.
• Weighing pan and anti-draft shield elements.
• Power adapter.
• User manual.
4.5. Cleaning
Balance should be cleaned with damp cloth by gentle wiping the smudges of
dirt. In order to clean the weighing pan of the balance, please remove it from
the rubber mandrels. Cleaning of the pan when assembled may cause
damage of the measuring system of the balance.
4.6. Plugging to mains
Balance can be connected to mains only with a power adapter that comes
standard with the balance. Nominal power supply of the power adapter
(specified on the data plate of the power adapter) should be compatible to
the power from mains. Plug the balance to mains – the plugging socket is
located at the back of balance’s housing. Balance display will light and show
name and number of software, after which balance will display 0.0000g. If
indication is different from zero, please press ZERO/TARE key.
4.7. Connecting peripheral devices
The balance must be unplugged from the mains before connecting or
disconnecting any peripheral devices (printer, PC computer). Use only
peripheral devices recommended by the manufacturer with your balance.
These have been ideally coordinated to your balance. On connecting a
peripheral device, plug the balance to mains.