9. WEIGHING
Before start of weighing process or in case of essential change of ambient
conditions at a workstation (e.g. ambient temperature change at a workstation more
than 3
o
C) the balance requires adjusting. the procedure of balance adjustment is
described further in this user manual.
• Before start of weighing procedure, it is recommended to load the balance’s
weighing pan a few times with mass close to balance max capacity,
• Check if unloaded balance indicates “precise zero” -
and whether
measurement is stable -
, if not press ESC/TARE key;
• Press Units key to set a measuring unit:
[g], [mg], [ct], and also if enabled in factory menu [kg], [lb], [oz], [ozt], [dwt],
[t], [mom], [gn],
• place weighed object on balance’s weighing pan and read result only on
stabilization of measurement result,
• mass indication of a load placed on balance’s weighing pan can be tarred for
multiple times by pressing ESC/TARE key ((pay attention not to exceed
maximal capacity of a balance by applying multiple tare function).
During times between carrying out the following measurement series do not unplug
the balance from mains. It is recommended to switch off balance’s display by
pressing ON/OFF key. On repeated pressing of the ON/OFF key the balance is
ready for operation and carrying out the following measurements.