49
12.4.2. External adjustment
The external adjustment in balances PS series should be carried out with
an external mass standard / weight class F
2
Start of external adjustment process
balances with internal adjustment balances with external adjustment
Balance will display a command ordering unloading of weighing pan
unloAd (weighing pan should be empty). When weighing pan is unloaded,
press PRINT key. The balance determines mass of empty pan. Place an
external adjustment weight with mass determined on balance’s display and
press PRINT key. On completing adjustment process the balance returns
to submenu P1.2 ECal.
Return to weighing
(see point 8.2. Return to weighing).
If a balance is verified, then user does not access to carrying out external
adjustment process.