Adjustment upon passage of specified time interval
The scale is equipped with RTC, due to this time of each completed
process is registered.
Adjustment process is triggered automatically after passage of particular
time interval, set in <1.6.CAC> parameter.
Adjustment process, when in progress, is signalled with display of 'dash',
< - >.
Upon completed adjustment process, the scale automatically returns
to the weighing mode.
12.5. Automatic Internal Adjustment Time
option available for non-verified scales exclusively
Parameter determining time interval between successive automatic internal
adjustments. The time interval is declared in hours and ranges between 0.5 [h]
and 12 [h].
Procedure:
Enter <P1.CAL / 1.6.CAC> submenu.
Press
key, available values, given in hours, are displayed
successively one by one:
Available values: 05 H, 1 H, 2 H, 3 H, 4 H, 5 H, 6 H, 7 H, 8 H, 9 H, 10 H, 11 H,
12 H.
12.6. Adjustment Test
Adjustment test function enables comparison of the internal adjustment results
with the value recorded in factory parameters. Such comparison is used for
determining drift of scale sensitivity over time.
Procedure:
Enter <P1.CAL / 1.4.AtS> submenu, internal adjustment process starts
automatically.
Adjustment process, when in progress, is signalled with display of <CAL>
text.
Upon completed adjustment process, <1.4.CAtS> parameter is displayed
automatically.