Chapter 6 MATH and Measurement RIGOL
MSO1000Z/DS1000Z User’s Guide 6-29
Voltage Parameters
Figure 6-5 Voltage Parameters
1. Vmax: the voltage value from the highest point of the waveform to the GND.
2. Vmin: the voltage value from the lowest point of the waveform to the GND.
3. Vpp: the voltage value from the highest point to the lowest point of the
waveform.
4. Vtop: the voltage value from the flat top of the waveform to the GND.
5. Vbase: the voltage value from the flat base of the waveform to the GND.
6. Vamp: the voltage value from the top of the waveform to the base of the
waveform.
7. Vupper: the actual voltage value corresponding to the threshold maximum
value.
8. Vmid: the actual voltage value corresponding to the threshold middle value.
9. Vlower: the actual voltage value corresponding to the threshold minimum
value.
10. Vavg: the arithmetic average value on the whole waveform or on the gating
area. The calculation formula is as follows:
i
x
Average =
n
∑
(6-2)
Wherein,
i
x
is the measurement result of the
i
th point,
n
is the number of
points being measured.
11. Vrms: the root mean square value on the whole waveform or the gating area.
The calculation formula is as follows.
Preshoot
Overshoot
Vamp
Vbase
Vmax
Vmin
Vpp
Vtop
Vupper
Vmid
Vlower
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
BrandRigol
ModelDS1104Z Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals