WARNING
To avoid electric shock when using the probe, please make sure
that the insulated wire of the probe is in good condition and do
not touch the metallic part of the probe when the probe is
connected to high voltage source.
Probe Compensation
When the probes are used for the first time, you should compensate the probes
to make them match the input channels of the oscilloscope. Non-compensated
or poorly compensated probes may cause measurement inaccuracy or errors.
The probe compensation procedures are as follows:
1. Perform Step 1, 2, 3 and 4 specified in "Function Inspection".
2. Check the displayed waveforms and compare them with the waveforms
shown in Figure 10.
Figure 10 Probe Compensation
3. Use the probe compensation adjustment tool provided in the accessories
to adjust the low-frequency compensation adjustment hole on the probe
until the displayed waveform is consistent with the "Perfectly
compensated" waveform shown in Figure 10.