RIGOL Chapter 2 Command System
2-190 MSO7000/DS7000 Programming Guide
Return
Format
The query returns GRE, LESS, GLES, or GOUT.
Example
:TRIGger:DELay:TYPE GOUT /*Sets the delay type to "><".*/
:TRIGger:DELay:TYPE? /*The query returns GOUT.*/
:TRIGger:DELay:TUPPer
Syntax
:TRIGger:DELay:TUPPer <time>
:TRIGger:DELay:TUPPer?
Description
Sets or queries the upper limit of delay time of the Delay trigger. The default unit is
s.
Parameter
Name
Type
Range
Default
<time>
Real
8.01 ns to 10 s
2 μs
Remarks
This command is available when the delay type is LESS, GOUT, or GLESs.
Return
Format
The query returns the upper limit of delay time in scientific notation.
Example
:TRIGger:DELay:TUPPer 0.002 /*Sets the upper limit of delay time to 2 ms.*/
:TRIGger:DELay:TUPPer? /*The query returns 2.000000E-3.*/
:TRIGger:DELay:TLOWer
Syntax
:TRIGger:DELay:TLOWer <time>
:TRIGger:DELay:TLOWer?
Description
Sets or queries the lower limit of delay time of the Delay trigger. The default unit is
s.
Parameter
Name
Type
Range
Default
<time>
Real
8 ns to 9.9 s
1 μs
Remarks
This command is available when the delay type is GREater, GOUT, or GLESs.
Return
Format
The query returns the lower limit of delay time in scientific notation.
Example
:TRIGger:DELay:TLOWer 0.002 /*Sets the lower limit of delay time to 2 ms.*/
:TRIGger:DELay:TLOWer? /*The query returns 2.000000E-3.*/
:TRIGger:DELay:ALEVel
Syntax
:TRIGger:DELay:ALEVel <level>
:TRIGger:DELay:ALEVel?
Description
Sets or queries the upper limit of the trigger level of Delay trigger. The unit is the same
as that of the current amplitude.
Parameter
Name
Type
Range
Default
<level>
Real
Lower limit to (5 × VerticalScale - OFFSet)
0 V
Remarks
For VerticalScale, refer to the :CHANnel<n>:SCALe command; for OFFSet, refer to
the :CHANnel<n>:OFFSet command.

Table of Contents

Other manuals for Rigol MSO7000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO7000 Series and is the answer not in the manual?

Rigol MSO7000 Series Specifications

General IconGeneral
BrandRigol
ModelMSO7000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals